Test & Measurement


Pulsed S-parameter measurements for GSM amplifiers

25 August 2004 Test & Measurement

Many devices, including the GSM power amplifiers used in cellular handsets, were not designed to operate continuously. The GSM system operates in the 900 or 1800 MHz bands and uses Gaussian minimum shift keying (GMSK), which is a type of digital FM modulation. The output power is measured during a burst, which is 577 µs long and repeats every 4,6 ms. Traditional measurement techniques are not adequate for this type of burst signal.

Therefore, pulsed-RF techniques must be employed during this burst period for accurate measurements of s-parameters.

This article provides an application guide to understanding general s-parameter measurements with a vector network analyser under pulsed-RF conditions for GSM power amplifiers. Most modern vector network analysers have a receiver and trigger functionality to make such measurements. A brief description of pulsed-RF measurements, the test set-up for a GSM amplifier and some s-parameter results are presented.

Pulsed RF signals

Most power amplifiers (PAs) are not designed to transmit with 100% duty cycle. If they were, the PA device would heat up, drawing considerable current. Transistors tested on a wafer would be subject to extreme heating, causing thermal stress. To avoid this problem, s-parameter measurements are often taken with the PA in a pulsed condition.

A pulsed RF signal (PRF) will have a frequency spectrum composed of a series of spectral lines with an envelope described by a sinc function (ie, sin(x)/x). The spacing of the lines ("1/T") is set by the PRF frequency while the envelope shape is fixed by the pulse width (assuming small rise and fall times to pulse width). The spacing of the spectral lines is equal to 1/period = 1/T as shown in Figure 1.

Figure 1. Spectrum of a PRF signal with spacing ‘1/T’ equal to one over the pulse period
Figure 1. Spectrum of a PRF signal with spacing ‘1/T’ equal to one over the pulse period

The GSM signal is a 577 µs pulse with a period of 4,6 ms, which equates to a PRF frequency of approximately 217 Hz. This spectrum's 'size' determines how much IF bandwidth (IFBW) is necessary in the vector network analyser (VNA) receiver for accurate PRF measurements. In this case, the entire spectrum can fit within an IFBW of 1 kHz:

There may be some outlying spectral components at the edges so some additional averaging may be needed to reduce the effects of the edge frequency components. The VNA must be aligned in time with the PRF; otherwise, all the transmitted s-parameter data points will not be collected properly. This is the fundamental reason for an externally 'triggered' signal to the VNA.

Triggered measurements

As shown by Equation 1, the received signal captured by the VNA fits entirely inside a typical IFBW (ie, 1 kHz), which can be set between 1 and 30 kHz in most VNAs. In the time domain, the sampling needs to occur during the amplifiers' 'on-period' of the pulse in order to capture the s-parameter data. This is achieved by externally triggering the VNA to measure the appropriate points in time. As shown in Figure 2, the external trigger pulse arrives at the VNA some time before the RF pulse in order to account for the 'latency' inside the VNA. 'Latency' is the term used to describe the delay from one transmission point to another. Latency occurs in many electronic systems where there is a propagation delay between source and receiver.

Figure 2. Timing of external trigger pulse to VNA and GSM RF burst
Figure 2. Timing of external trigger pulse to VNA and GSM RF burst

In this case, the VNA requires a delay between its received external trigger signal and the Tx-Enable signal. According to an Anritsu application note, the delay should be on the order of 50 to 100 µs although starting later is allowed. The sampling can begin sometime after the RF pulse has settled unless the initial data points are of additional interest. The sampling can continue for a substantial portion of the pulse but should not continue beyond the end of the pulse.

As a general rule, the IFBW must be greater than 1/pulse width to keep sampling from overrunning the pulse (IFBW = 10 kHz with averaging turned on is acceptable on the Anritsu MS4623B VNA). Due to pulse settling time, VNA internal filtering and latency issues, some safety margin is required. This will vary greatly depending on set-up and may require some experimentation. If the IFBW is too small, the sampled s-parameters will contain some noise as a result of not completely capturing the full GSM burst.

Measurement set-up

In order to validate the theory, the RF Micro Devices RF3110 quad-band power amplifier will be measured and S11 and S21 s-parameters captured on the Agilent MS4623B vector network analyser (VNA) using an Anritsu Display Capture program.

The test set-up is shown in Figure 3. It includes an arbitrary waveform generator (AWG 320) for the ramp profile and a 2,5 Vp-p square wave trigger signal, which was used to synchronise the VNA with the Transmit Enable (Tx-En) signal on the RF3110 (see Figure 4).

Figure 3. Set-up for making PRF measurements and capturing S-parameters for the RF3110 power amplifier
Figure 3. Set-up for making PRF measurements and capturing S-parameters for the RF3110 power amplifier

Figure 4. Tx-enable signal (2 V pulse) showing the ‘leading’ external trigger signal (2,5 Vp-p square wave)
Figure 4. Tx-enable signal (2 V pulse) showing the ‘leading’ external trigger signal (2,5 Vp-p square wave)

In order to achieve a 'leading' external trigger signal, the phase angle of the trigger signal was then changed to allow offset waveforms. The trigger signal should lead the Tx-Enable signal by 50 to 100 µs. A lead time of 70 µs corresponds to a phase setting of 32 degrees on the AFG 320 arbitrary waveform generator.

A calibrated input signal of -10 dBm was applied to the input port of the RF3110 evaluation board (EVB). A 20 dB attenuator was also placed at the output port to avoid saturating the VNA front end, which accepts a received signal of +5 dBm. DC voltages were also sequentially applied to Vreg (2,8 V) and Band-Select (2 V). A Vramp waveform of 1,4 V that conforms to the ETSI timing mask was then applied to the RF3110 in order to take the full-power s-parameter measurements.

S-parameter measurements

Small-signal measurements of S21 and S11 were taken on the RF3110. Associated plots are shown in Figures 5, 6 and 7. S21 represents the actual gain of the device and S11 reveals the input impedance when the device is turning off and on. Without using triggered measurements both of these parameters would not be accurately reported. S22 data is not collected since it represents the output impedance of the RF3110, which is meaningless for impedance matching to the antenna.

Figure 5. S21 gain curves of RF3110: Externally triggered VNA (flat top line with marker ‘3’) and non-pulsed (internally triggered VNA; narrow-spiked waveform). The ‘external trigger rate too fast’ message is superficial and only occurs with the VNA ‘screen capture’ operation
Figure 5. S21 gain curves of RF3110: Externally triggered VNA (flat top line with marker ‘3’) and non-pulsed (internally triggered VNA; narrow-spiked waveform). The ‘external trigger rate too fast’ message is superficial and only occurs with the VNA ‘screen capture’ operation

Figure 6. S21 gain curves of RF3110 from 800 to 4000 MHz: externally triggered VNA (flat top line with marker ‘3’) and non-pulsed (internally triggered VNA; narrow-spiked waveform)
Figure 6. S21 gain curves of RF3110 from 800 to 4000 MHz: externally triggered VNA (flat top line with marker ‘3’) and non-pulsed (internally triggered VNA; narrow-spiked waveform)

Figure 7. Input reflection coefficient, S11, of RF3110 from 800 to 4000 MHz: externally triggered VNA (orange curve) and non-pulsed (internally triggered VNA; green curve)
Figure 7. Input reflection coefficient, S11, of RF3110 from 800 to 4000 MHz: externally triggered VNA (orange curve) and non-pulsed (internally triggered VNA; green curve)

Figure 5 shows the gain plot for the external and internal triggered measurement. The top line is the gain curve for the VNA externally triggered and its markers are shown to the right (ie, 32 dB of gain). The default internally triggered signal shows spikes, indicating that the VNA is not capturing the gain (S21) of the transmitted GSM signal correctly. The 'external trigger rate too fast' message is superficial and only occurs with the VNA screen capture operation.

Figure 6 shows the same plot over larger frequency band.

The effects of externally triggering the GSM amplifier for the input reflection coefficient, S11, is shown in Figure 7.

Conclusion

A technique for pulsed S-parameter measurements on GSM amplifiers has been presented along with some details on how to make successful measurements using a network analyser. Plots of S11 and S21 for a pulsed RF signal with an external and internal trigger are displayed to show the results of using the external trigger on the VNA for PRF environments. It is the hope of the author that this information may be beneficial to those taking small-signal measurements on devices with pulsed RF.



Credit(s)



Share this article:
Share via emailShare via LinkedInPrint this page

Further reading:

New GNSS passive patch antenna
RF Design Telecoms, Datacoms, Wireless, IoT
The HP24510A from Taoglas is a stacked patch GNSS passive antenna that operates from 1215 to 1610 MHz covering the L1/L2 GNSS spectrum.

Read more...
InterCal introduces a new range of calibrators
Intercal Test & Measurement
Calibration is a critical process in numerous industries, from manufacturing to healthcare, ensuring that instruments and devices provide accurate and reliable measurements.

Read more...
Power Energy Meter for solar projects
Mimic Components Test & Measurement
The Mi550 Power Energy Meter is a handheld three-phase power quality analyser that connects externally with Rogowski coils or voltage-type CTs, allowing for testing without disconnection.

Read more...
MT8870A for NTN NB-IoT testing
Tamashi Technology Investments Test & Measurement
Samsung Electronics has selected the Anritsu Universal Wireless Test Set MT8870A for use on the mass-production lines of the Galaxy S25.

Read more...
XJLink-PF40 JTAG controller
ASIC Design Services Test & Measurement
XJTAG, a specialist in electronic testing, has released its new XJLink-PF40 JTAG controller together with version 4 of its popular PCB software testing suite.

Read more...
INTEGRA Biosciences’ reagent dispenser now supports SiLA-2 integration
Test & Measurement
This new functionality will make it even easier to integrate the WELLJET into automated laboratory workflows, helping labs to enhance their productivity and reproducibility.

Read more...
6 – 18 GHz driver amplifier
RF Design Telecoms, Datacoms, Wireless, IoT
The QPA0022D from Qorvo is a high-performance driver amplifier covering a range of 6 to 18 GHz and fabricated on Qorvo’s production 0,15 µm pHEMT process.

Read more...
Advanced portable spectrum analyser
Vepac Electronics Test & Measurement
HAROGIC has introduced the SAE-200, a USB-based real-time spectrum analyser that operates from 9 to 20 GHz with a 100 MHz bandwidth.

Read more...
Water detection sensor
Test & Measurement
Recently, a major global automaker experienced significant issues caused by coolant leaking within the battery system, resulting in short circuit and a subsequent fire that destroyed the vehicle.

Read more...
New atom-based thermometer measures temperature more accurately
Test & Measurement
Beyond precision science, the new thermometer could have wide-ranging applications in challenging environments from spacecraft to advanced manufacturing plants, where sensitive temperature readings are essential.

Read more...